Command reference
R&S
®
ZNL/ZNLE
910User Manual 1178.5966.02 ─ 20
Enable the MCTYpes mode before defining the calibrations to be performed using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine. Disable the MCTYpes
mode after the calibrations were saved using [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected[:DUMMy]. The latter must be called only once, with <Ch>
referring to one of the channels to be calibrated (see the example below).
Parameters:
<Boolean>
Example: Suppose that there are 3 channels in the current channel setup
and you want to calibrate channels 1 and 2. For channel 1, you
want to perform a Reflection Normalization (Short) on ports 1
and 2, for channel 2 an OSM calibration on port 2. This calibra-
tion requires measuring standards as shown in the table below.
Proceed as follows:
:CORRection:COLLect:CHANnels:MCTYpes ON
Activate the MCTYPes mode.
:SENSe1:CORRection:COLLect:METHod:DEFine
'Parser Test SFK Ch1', RSHort, 1, 2
:SENSe2:CORRection:COLLect:METHod:DEFine
'Parser Test SFK Ch2', FOPort, 2
Declare the calibrations to be performed. Then connect the
Short standard to port 1
:SENSe:CORRection:COLLect:ACQuire:SELected
SHORt, 1
Acquire calibration data for the short on port 1 (needed for cali-
bration in channel 1). Then connect the Open standard to port 2.
:SENSe:CORRection:COLLect:ACQuire:SELected
OPEN, 2
Acquire calibration data for the open on port 2 (needed for cali-
bration in channel 2). Then connect the Short standard to port 2.
:SENSe:CORRection:COLLect:ACQuire:SELected
SHORt, 2
Acquire calibration data for short on port 2 (needed for calibra-
tions in channel 1 and 2). Then connect the Match standard to
port 2.
:SENSe:CORRection:COLLect:ACQuire:SELected
MATCh, 2
Acquire calibration data for match on port 2 (needed for calibra-
tion in channel 2).
:SENSe1:CORRection:COLLect:SAVE:SELected
Complete all calibrations. This command is required only once
and has to address one of the calibrated channels (channel 1 or
2 in this example).
:CORRection:COLLect:CHANnels:MCTYpes OFF
Deactivate the MCTYPes mode. Recommended for compatibility
with manual operation.