Command reference
R&S
®
ZNL/ZNLE
935User Manual 1178.5966.02 ─ 20
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:STATe <Boolean>
Defines whether the second reference frequency f
2
is used for the offset loss approxi-
mation.
The second reference frequency and its associated loss value can be defined using
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:FREQuency and
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> OFF (0)
f
2
not used (default)
ON (1)
f
2
used
Manual operation: See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 534
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe] <Boolean>
Toggles length/loss/fixture compensation for physical port <PhyPt> ON/OFF.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> ON (1): compensation active
OFF (0): compensation inactive
Manual operation: See "Active" on page 534
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Returns whether a direct fixture compensation has been carried out at port no.
<PhyPt>. A direct fixture compensation resets the offset parameters to zero, the ana-
lyzer uses calculated transmission factors instead.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer.
Return values:
<Boolean> 1 - direct fixture compensation data used
0 - no direct fixture compensation data used