Example:
*RST; CORR:OFFS:DFC?
Reset the instrument and query whether the analyzer uses direct
fixture compensation data at port 1. The response is 0.
Usage: Query only
Manual operation: See "Direct Compensation" on page 546
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe] <Boolean>
Resets the offset parameters for all test ports to zero and the reference frequency to 1
GHz or queries whether any of the offset parameters are different from zero.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer. This numeric suffix is ignored; the
command affects the parameters of all ports.
Parameters:
<Boolean> The parameter function depends on whether the command is
used as a setting command or as a query: For setting command:
ON - no effect
OFF - resets all length offsets to zero and the reference fre-
quency to 1 GHz
For query:
1 - at least one length offset parameter is different from its
default value
0 - all length offsets are zero / set to default
*RST: OFF
Example:
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset param-
eters have been reset as well. The response is 0.
Manual operation: See "Reset Offsets" on page 535
[SENSe<Ch>:]CORRection:PCAL <ActivatePowerCals>
Activates or deactivates all power calibrations.
Suffix:
<Ch>
.
Calibrated channel number
Setting parameters:
<ActivatePowerCals> NONE | ALL
NONE deactivate all power calibrations.
ALL activate all power calibrations.
*RST: n/a