Command reference
R&S
®
ZNL/ZNLE
961User Manual 1178.5966.02 ─ 20
[SENSe<Ch>:]SEGMent:POWer[:LEVel]:CONTrol <Boolean>
Selects common or independent internal source "Power" settings for the sweep seg-
ments.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON use independent power settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].
OFF reset the power in all sweep segments to the power for
unsegmented sweeps, defined via SOURce<Ch>:
POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]. ON
will not restore the previous values.
The parameter is automatically switched to ON when a segment
power is entered using [SENSe<Ch>:]SEGMent<Seg>:
POWer[:LEVel].
*RST: OFF
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus -10 dBm internal source power.
SEGM:POW -20
Decrease the power to -20 dBm.
SEGM:POW:CONT OFF
Couple the powers in all segments and reset the power in seg-
ment no. 1 to the initial value.
Manual operation: See "Optional Columns" on page 337
[SENSe<Ch>:]SEGMent<Seg>[:STATe] <Boolean>
Activates or deactivates the sweep segment <Seg>. Sweep points belonging to inac-
tive segments only are not measured.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<Boolean> ON | OFF - activates or deactivates the measurement in sweep
segment <Seg>.
*RST: ON
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM OFF
Disable the measurement in the created sweep segment.
Manual operation: See "Table Columns" on page 334