Parameters:
<Frequency> Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range: Stimulus values outside the frequency range of the
analyzer are allowed.
*RST: n/a
Default unit: Hz
<TransCoeff> Transmission coefficient
Range: -300 dB to +200 dB
*RST: n/a
Default unit: dB
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Example:
See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 983
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 515
SOURce:POWer:CORRection:TCOefficient[:STATe] <Boolean>[, <DeEmbedding
Position>]
Enables or disables the use of two-port transmission coefficients for power correction.
Parameters:
<Boolean> Enables (ON) or disables (OFF) the two-port transmission coeffi-
cients.
*RST: n/a
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
VNA command reference