R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Programming Examples
Condensed Programming Examples
Operating Manual 1145.1084.12 – 30 1115
// C:\Program Files\Rohde&Schwarz\Network Analyzer\Calibration\Data
// the file name in the commands must not contain the path !
:MMEMORY:STORE:CORRection 1, 'OSM1 TOSM12.cal'
//
// load cal file from calibration file pool
:MMEMORY:LOAD:CORRection 1, 'OSM1 TOSM12.cal'
Create Calibration Kit
The following example defines the properties of different calibration standards.
// Reset the analyzer
*RST
:SYSTEM:DISPLAY:UPDATE ON
////////////////////////////////////////////////////////////////////////////////
//
// Remarks: if a cal kit name contains "Ohm", this "Ohm" (case insensitive) is replaced by the
// capital greek omega for the cal kit name used in the DataBase !!!
//
////////////////////////////////////////////////////////////////////////////////
// One port standards
////////////////////////////////////////////////////////////////////////////////
//
// All commands to set the parameters of a physical standard
// have the following first 4 parameters :
// standard label, serial number, min frequency, max frequency
//
// The one port standard REFLECT and the two port standard SYMMETRIC NETWORK
// have 2 possibilities for parameter 10:
// OPEN: parameters 6 ... 9 are polynomial coefficients of the capacitance
// SHORt: parameters 6 ... 9 are polynomial coefficients of the inductance
// Define OPEN (female): + length, loss, capacitance[4]
:SENSE:CORRECTION:CKIT:N50:FOPEN "Parser N50 Ohm","Ser 1-
1",1.1MHZ,2.1MHZ,0.1m,11,12,13,14,15
// Set active cal kit for the connector N50. This active cal kit is used in the