Simple Measurement Tasks
Operating Manual 1145.1084.12 – 30 69
To load a setup stored on a USB memory stick...
1. Plug a USB stick into any of the USB type A connectors on the front or rear panel of the analyzer.
2. Click File — Open to open the Open dialog.
3. Open the Look in: drop-down list and select the network drive labeled Removable Disk (most
likely, drive E:) as file location.
The setup files are displayed in the center of the dialog.
4. Select a file and click Open to load the setup as the active setup.
Setting up a Sweep
Configurations of the different sweep types require different procedures. The following examples list the
essential settings for various sweep types.
To Set up a Lin. Frequency Sweep...
1. Press the Sweep key in the Channel keypad or click Channel - Sweep to access the Sweep
submenu.
2. Click Sweep Type – Lin. Frequency.
3. Click Channel – Stimulus to access the submenu defining the sweep range.
4. To define a sweep range confined by a start and a stop frequency, use Start and Stop.
Alternatively, to define a sweep range of definite width around a known center frequency (e.g. an
expected peak), use Center and Span.
5. To vary the number of measurement points (e.g. to improve the resolution of the sweep or reduce
the sweep time), click Channel – Sweep – Number of Points.
If the active diagram is a Cartesian diagram, the x-axis is linearly scaled and labeled with the Start and
Stop frequencies.
To Set up a Log. Frequency Sweep...
1. Press the Sweep key in the Channel keypad or click Channel - Sweep to access the Sweep
submenu.
2. Click Sweep Type - Log. Frequency.
3. Click Channel - Stimulus to access the submenu defining the sweep range.
4. To define a sweep range confined by a start and a stop frequency, use Start and Stop.
Alternatively, to define a sweep range of definite width around a known center frequency (e.g. an
expected peak), use Center and Span.
5. To vary the number of measurement points (e.g. to improve the resolution of the sweep or reduce
the sweep time), click Channel - Sweep - Number of Points.
If the active diagram is a Cartesian diagram, the x-axis is logarithmically scaled and labeled with the Start
and Stop frequencies.