R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 809
Confirmed, command or query.
The query returns 1 as soon as the executed program is stopped or
paused.
See PROGram[:SELected]:EXECute or PROGram[:SELected]:RETVal.
PROGram[:SELected]:RETVal?
Queries the return value of an application or process started via PROGram[:SELected]:EXECute.
Since PROG:RETVal? is a non-blocking command, it has to be preceded by a PROG:WAIT (command or
query).
PROG:EXEC 'iecwin32.exe'
Run iecwin32.exe.
PROG:WAIT
Lock command execution until iecwin32.exe is stopped.
PROG:RETVal?
Query the return value (0 in case of success)
ROUTe<Ch>...
This subsystem configures the internal signal paths of a connected switch matrix R&S ZV-Z81/-82 or a
multiport test set R&S ZV-83. For details refer to System Configuration – Matrix Configuration in the GUI
reference chapter.
ROUTe<Ch>:CFILe <Boolean>
Overwrites the current settings of the switch matrix/multiport test set with the settings stored in the
configuration file matrix.mtx. The file is stored in the directory C:\Program Files\Rohde&Schwarz\Network
Analyzer\resources\extdev.
Channel number. This parameter is ignored; the settings in the configuration
file affect all channels no. Ch1 to Ch9.
ON - The multiport test set uses the default settings of the configuration file
OFF - The default settings have been changed (for query only)