R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 610
SCPI Command Reference
ABORt...
This subsystem resets the trigger system and places all trigger sequences in the IDLE state. Any actions
related to the trigger system that are in progress, such as a sweep or acquiring a measurement, shall also
be aborted as quickly as possible.
ABORt
Aborts a running measurement and resets the trigger system.
In continuous sweep mode (INITiate<Ch>:CONTinuous ON) the ZVA automatically starts a new
sweep.
INIT:CONT OFF
Set the analyzer to "Single Sweep" mode.
INIT:IMM
Start a new single sweep sequence.
ABORt
Abort the running measurement.
INIT:IMM
Initiate new sweep ("Restart").
CALCulate
CALCulate<Chn>:CLIMits...
This subsystem controls the composite limit check.
CALCulate<Chn>:CLIMits:FAIL?
Returns a 0 or 1, to indicate whether or not a global, composite limit check on several traces has failed.
In remote control, the limit check result is calculated once at the end of each sweep. If the limit lines
are changed, a new sweep is required to obtain updated limit check results. In single sweep mode
(INITiate<Ch>:CONTinuous OFF), the new sweep must be started explicitly using
INITiate<Ch>[:IMMediate]:SCOPe ALL | SINGle. This behavior is different from manual control