SCPI Command Reference
Operating Manual 1145.1084.12 – 30 811
[SENSe...]
[SENSe<Ch>:]AVERage...
This subsystem sets sweep averaging parameters. The sweep average is a noise-reduction technique
which consists of calculating each measurement point as an average of the same measurement point over
several consecutive sweeps.
In contrast to the sweep count (for single sweep mode, [SENSe<Ch>:]SWEep:COUNt), averaging is
always channel-specific. Both features are completely independent from each other.
[SENSe<Ch>:]AVERage:CLEar
Starts a new average cycle, clearing all previous results and thus eliminating their effect on the new cycle.
Channel number. If unspecified the numeric suffix is set to 1.
SENS1:AVER:COUN 15; :AVER ON
Set the average factor for channel 1 to 15 (the mnemonic SENS1 can be
omitted) and enable the sweep average.
SENS1:AVER:COUN:CURR?
Query the currently measured sweep.
AVER:COUN 5; CLE
Reduce the average factor and restart the average.
[SENSe<Ch>:]AVERage:COUNt <factor>
Defines the number of consecutive sweeps to be combined for the sweep average (Average Factor).
Confirmed, command or query
See [SENSe<Ch>:]AVERage:CLEar
[SENSe<Ch>:]AVERage:COUNt:CURRent?
Queries the number of the sweep which is currently measured. Use this command to monitor the progress