R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 962
[SENSe<Ch>:]SWEep:COUNt <No_of_Sweeps>
Defines the number of sweeps to be measured in single sweep mode (INITiate<Ch>:CONTinuous
OFF).
Channel number. If the single sweep mode settings apply to all channels
(INITiate<Ch>[:IMMediate]:SCOPe ALL), this suffix is ignored; see
program example in section Single Sweep Mode.
Number of consecutive sweeps to be measured.
Confirmed, command or query
See CALCulate<Chn>:DATA:NSWeep:FIRSt?.
[SENSe<Ch>:]SWEep:DETector:TIME <det_time>
Defines the detector observation time for ratios and wave quantities if a Peak or RMS detector is active.
The detector is selected together with the measured quantity (CALCulate<Ch>:PARameter:MEASure
or CALCulate<Ch>:PARameter:SDEFine).
Confirmed, command or query
*RST; :CALC:PAR:MEAS 'TRC1', 'a1D1RMS'
Select the wave quantity a1 for the default trace and activate the RMS
detector.
SWE:DET:TIME 1
Specify an observation time of 1 s at each sweep point.
[SENSe<Ch>:]SWEep:DWELl <delay>
Defines the Meas. Delay for each partial measurement. Setting a delay disables the automatic calculation
of the (minimum) sweep time; see [SENSe<Ch>:]SWEep:TIME:AUTO.
Meas. delay before each partial measurement.
0 s to 3.179551E+002 s. [s] Changing the delay leaves the number of points
unchanged but has an impact on the duration of the sweep
([SENSe<Ch>:]SWEep:TIME).