Operating Manual 1145.1084.12 – 30 966
SWE:TIME 2
Increase the total sweep time to 2 s.
SWE:DWEL?
Query the meas. delay for each partial measurement again. The delay is increased
by 1 s divided by the total number of partial measurements per sweep.
[SENSe<Ch>:]SWEep:TIME:AUTO <Boolean>
When enabled, the (minimum) sweep duration is calculated internally using the other channel settings and
zero delay ([SENSe<Ch>:]SWEep:DWELl).
Note: The automatically calculated sweep duration is ignored for the sweep types Time and CW Mode.
ON | OFF - Turns the automatic calculation of the sweep time on or off. OFF is also
set if the sweep duration or delay is set explicitly using [SENSe<Ch>:]SWEep:TIME
or [SENSe<Ch>:]SWEep:DWELl.
Confirmed, command or query
SWE:TIME 1
Set a total sweep time of 1 s.
SWE:TIME:AUTO?
A query returns the value 1.
[SENSe<Ch>:]SWEep:TYPE LINear | LOGarithmic | SEGMent | POWer | CW | POINt
| PULSe | IAMPlitude | IPHase
Selects the sweep type, i.e. the sweep variable (frequency/power/time) and the position of the sweep
points across the sweep range.
Lin. frequency sweep at constant source power
(SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate][:AMPlitude]). The
stimulus frequency ([SENSe<Ch>:]FREQuency:...) is swept in equidistant steps
over the frequency range. In a Cartesian diagram, the x-axis is a linear frequency
axis.
Log. frequency sweep. The frequency is swept in equidistant steps on a logarithmic
scale. In a Cartesian diagram, the x-axis is a logarithmic frequency axis.
Segmented frequency sweep. The sweep range is composed of several continuous
frequency ranges or single frequency points defined by means of the commands in
the [SENSe<Ch>:]SEGMent<Seg>... subsystem.
Power sweep. The measurement is performed at constant frequency
(SOURce<Ch>:FREQuency<Pt>:CW|FIXed) but with variable generator power that