R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Programming Examples
Condensed Programming Examples
Operating Manual 1145.1084.12 – 30 1112
// Loop over these 2 commands
:INITIATE:IMMEDIATE; *WAI
:TRACe:COPY:MATH 'LastExtr', 'Trc1'
:INITIATE:IMMEDIATE; *WAI
:TRACe:COPY:MATH 'LastExtr', 'Trc1'
// ..............
// Continuous sweep mode
:INITIATE:CONTINUOUS ON
//
Retrieving the Results of Previous Sweeps
The commands CALCulate<Ch>:DATA:NSWeep...? SDATa, <Trace_Hist_Count> retrieve the
results of any sweep within a previously defined single sweep group. This means that, in single sweep
mode, you can first measure a specified number of sweeps (SENSe<Ch>:SWEep:COUNt <sweeps>) and
then read any of the data traces acquired.
This feature has no equivalent in manual control where always the last data trace is displayed.
// Reset the analyzer
*RST
:SYSTEM:DISPLAY:UPDATE ON
// Create a second channel with a second trace.
:CALCULATE2:PARAMETER:SDEFINE "Trc2","S11"
:CALCULATE2:PARAMETER:SELECT "Trc2"
:DISPLAY:WINDOW2:STATE ON
:DISPLAY:WINDOW2:TRACE1:FEED 'Trc2'
// Select active trace for the created channel 2. Adjust the number of sweep points.
:CALCULATE2:PARAMETER:SELECT "Trc2"
:SENSE1:SWEEP:POINTS 3
:SENSE2:SWEEP:POINTS 4
// Set sweep time and sweep count for the channels
// (3 traces per single sweep in channel 1, 4 traces in channel 2)
:SENSE1:SWEEP:TIME 1 S
:SENSE2:SWEEP:TIME 1 S
:SENSE1:SWEEP:COUNT 3
:SENSE2:SWEEP:COUNT 4
// State that INITIATE<Ch>:IMMEDIATE will start a single sweep for the