R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 812
of sweep averaging.
1 to the average factor ([SENSe<Ch>:]AVERage:COUNt) [1]
Device-specific, query only
See [SENSe<Ch>:]AVERage:CLEar
[SENSe<Ch>:]AVERage[:STATe] <Boolean>
Enable or disable the sweep average.
ON | OFF - Enables or disables the automatic calculation of the sweep average
over the specified number of sweeps ([SENSe<Ch>:]AVERage:COUNt).
Confirmed, command or query
See [SENSe<Ch>:]AVERage:CLEar
[SENSe<Ch>:]BANDwidth...
This subsystem sets the bandwidth of the IF measurement filter (resolution bandwidth).The forms
BANDwidth and BWIDth are equivalent.
[SENSe<Ch>:]BANDwidth|BWIDth[:RESolution] <bandwidth>
Defines the resolution bandwidth of the analyzer (Meas. Bandwidth).
Channel number. If unspecified the numeric suffix is set to 1.
1 Hz to 1 MHz (5 (10) MHz with option R&S ZVA-K17, 10 (30) MHz with option
R&S ZVA-K7; see Pulsed Measurements) [Hz]. UP and DOWN
increment/decrement the bandwidth in 1-2-5 steps for each decade. The analyzer
rounds up any entered value between these steps and rounds down values
exceeding the maximum bandwidth.
Confirmed, command or query