CALCulate<Chn>:LDEViation:CONStant <Constant>
Defines the constant value for the linearity deviation calculation.
Channel number used to identify the active trace.
The range depends on the trace format. For the default trace:
–200 dB to + 200 dB [dB]. The increment (UP, DOWN) is 10
-2
dB.
Device-specific, command or query.
See CALCulate<Chn>:LDEViation:AUTO
CALCulate<Chn>:LDEViation:ELENgth <El Length>
Defines the electrical length for the linearity deviation calculation. This command is available if the active
trace format is Phase or Unwrapped Phase, or if a mixer delay is measured.
Channel number used to identify the active trace.
0 m to + 1000 m [m]. The increment (UP, DOWN) is 10
-4
m.
Device-specific, command or query.
See CALCulate<Chn>:LDEViation:AUTO
CALCulate<Chn>:LDEViation:MODE ON | OFF | TRACking
Applies /discards the correction factors or re-calculates them for each trace (Tracking).
Channel number used to identify the active trace.
Apply the correction using the latest correction factors.
Repeat calculation for each measured trace and apply the correction.
Device-specific, command or query.
See CALCulate<Chn>:LDEViation:AUTO