Remote Control
R&S
®
FSVA/FSV
610Operating Manual 1176.7510.02 ─ 09
Return values:
<Options> The query returns a list of all installed and activated options,
separated by commas, where:
B<number> describes hardware options
K<number> describes software options
For PSA89600 emulation, the option is indicated as "B7J" for the
*OPT? query ("B7J, 140" if SYST:PSA:WID is activated).
(See SYSTem:PSA:WIDeband on page 971.)
Note that B99 indicates the R&S FSVA models (as opposed to
the R&S FSV models).
Example: B4,B5,B6,B7,B8,B10,B22,B30,B31,K7,K9
Usage: Query only
Manual operation: See "Versions+Options" on page 190
*PCB <Address>
Pass control back
Indicates the controller address to which remote control is returned after termination of
the triggered action.
Setting parameters:
<Address> Range: 0 to 30
Usage: Setting only
*PRE <Value>
Parallel poll register enable
Sets parallel poll enable register to the indicated value. The query returns the contents
of the parallel poll enable register in decimal form.
Parameters:
<Value> Range: 0 to 255
*PSC <Action>
Power on status clear
Determines whether the contents of the ENABle registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Remote Control – Commands