Remote Control
R&S
®
FSVA/FSV
645Operating Manual 1176.7510.02 ─ 09
Example:
CALC:LIM:ACP:ACH 30DB, 30DB
Sets the relative limit value for the power in the lower and upper
adjacent channel to 30 dB below the channel power.
CALC:LIM:ACP:ACH:ABS -35DBM, -35DBM
Sets the absolute limit value for the power in the lower and
upper adjacent channel to -35 dB.
CALC:LIM:ACP ON
Switches on globally the limit check for the channel/adjacent
channel measurement.
CALC:LIM:ACP:ACH:STAT ON
Switches on the limit check for the adjacent channels.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ACH:RES?
Queries the limit check result in the adjacent channels.
Manual operation:
See "Limit Checking" on page 329
CALCulate<n>:LIMit<k>:ACPower:ALTernate<channel>[:RELative] <LowerLimit>,
<UpperLimit>
This command defines the limit for the alternate adjacent channels for adjacent chan-
nel power measurements. The reference value for the relative limit value is the mea-
sured channel power.
Note that the relative limit value has no effect on the limit check as soon as it is below
the absolute limit defined with CALCulate<n>:LIMit<k>:ACPower:
ALTernate<Channel>:ABSolute. This mechanism allows automatic checking of
the absolute basic values of adjacent-channel power as defined in mobile radio stand-
ards.
Suffix:
<n>
.
Selects the measurement window.
<k> irrelevant
<Channel> 1...11
the alternate channel
Parameters:
<LowerLimit>,
<UpperLimit>
first value: 0 to 100dB; limit for the lower and the upper alternate
adjacent channel
*RST: 0 DB
Example:
CALC:LIM:ACP:ALT2 30DB, 30DB
Sets the relative limit value for the power in the lower and upper
second alternate adjacent channel to 30 dB below the channel
power.
Manual operation: See "Limit Checking" on page 329
Remote Control – Commands