Remote Control
R&S
®
FSVA/FSV
644Operating Manual 1176.7510.02 ─ 09
The result can be queried with CALCulate<n>:LIMit<k>:ACPower:ACHannel:
RESult. It should be noted that a complete measurement must be performed between
switching on the limit check and the result query, since otherwise no correct results are
available.
Suffix:
<n>
.
Selects the measurement window.
<k> irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM:ACP:ACH 30DB, 30DB
Sets the relative limit value for the power in the lower and upper
adjacent channel to 30 dB below the channel power.
CALC:LIM:ACP:ACH:ABS -35DBM, -35DBM
Sets the absolute limit value for the power in the lower and
upper adjacent channel to -35 dBm.
CALC:LIM:ACP ON
Switches on globally the limit check for the channel/adjacent-
channel measurement.
CALC:LIM:ACP:ACH:REL:STAT ON
Switches on the check of the relative limit values for adjacent
channels.
CALC:LIM:ACP:ACH:ABS:STAT ON
Switches on the check of absolute limit values for the adjacent
channels.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ACH:RES?
Queries the limit check result in the adjacent channels.
Manual operation:
See "Absolute Limit" on page 330
CALCulate<n>:LIMit<k>:ACPower:ACHannel:RESult
This command queries the result of the limit check for the upper/lower adjacent chan-
nel when adjacent channel power measurement is performed.
If the power measurement of the adjacent channel is switched off, the command pro-
duces a query error.
Suffix:
<n>
.
Selects the measurement window.
<k> irrelevant
Return values:
Result The result is returned in the form <result>, <result> where
<result> = PASSED | FAILED, and where the first returned value
denotes the lower, the second denotes the upper adjacent chan-
nel.
Remote Control – Commands