Remote Control
R&S
®
FSVA/FSV
734Operating Manual 1176.7510.02 ─ 09
Parameters:
<Frames> The maximum number of frames depends on
the .CALCulate<n>:SGRam:HDEPth on page 734
Range: 1 to depends on history depth
Increment: 1
*RST: 1
Example:
INIT:CONT OFF
Selects single sweep mode.
CALC:SGR:FRAM:COUN 200
Sets the number of frames to 200.
Manual operation: See "Frame Count" on page 250
CALCulate<n>:SGRam:FRAMe:SELect <Frame>
This command selects a specific frame for further analysis. The command is available
only if no measurement is running or after a single sweep has ended.
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<Frame> <frame_number> (if time stamp is off)
Selects the frame. The range is {0...number of recorded
frames-1}
<time> (if time stamp is on)
Time distance in seconds. It selects the frame that is x seconds
away from frame 0.
Example:
INIT:CONT OFF
Stop the continuous sweep.
CALC:SGR:FRAM:SEL -25
Selects frame number -25.
Manual operation: See "Select Frame" on page 250
CALCulate<n>:SGRam:HDEPth <HistoryDepth>
This command sets the number of frames to be stored in the R&S FSVA/FSV's mem-
ory.
Suffix:
<n>
.
1...4
Selects the measurement window.
Remote Control – Commands