Remote Control
R&S
®
FSVA/FSV
740Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:STATistics:SCALe:X:RANGe <Value>
This command defines the level range for the x-axis of the measurement diagram. The
setting is identical to the level range setting defined with the DISPlay[:
WINDow<n>]:TRACe<t>:Y[:SCALe] command.
Suffix:
<n>
.
irrelevant
Parameters:
<Value> 10dB to 200dB
*RST: 100dB
Example:
CALC:STAT:SCAL:X:RANG 20dB
Manual operation: See "x-Axis Range" on page 391
CALCulate<n>:STATistics:SCALe:X:RLEVel <Value>
This command defines the reference level for the x-axis of the measurement diagram.
The setting is identical to the reference level setting using the DISPlay[:
WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel command.
With the reference level offset <> 0 the indicated value range of the reference level is
modified by the offset.
The unit depends on the setting performed with CALCulate<n>:UNIT:POWer.
Suffix:
<n>
.
irrelevant
Parameters:
<Value> -120dBm to 20dBm
*RST: -20dBm
Example:
CALC:STAT:SCAL:X:RLEV -60dBm
Manual operation: See "x-Axis Ref Level" on page 391
CALCulate<n>:STATistics:SCALe:Y:LOWer <Value>
This command defines the lower limit for the y-axis of the diagram in statistical mea-
surements. Since probabilities are specified on the y-axis, the entered numeric values
are dimensionless.
Suffix:
<n>
.
selects the screen
Parameters:
<Value> 1E-9 to 0.1
*RST: 1E-6
Example:
CALC:STAT:SCAL:Y:LOW 0.001
Manual operation: See "y-Axis Min Value" on page 393