Remote Control
R&S
®
FSVA/FSV
770Operating Manual 1176.7510.02 ─ 09
MMEMory:STORe<n>:LIST <FileName>
This command stores the current list evaluation results in a <file name>.dat file.
The file consists of a data section containing the list evaluation results.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName> <file name>
Example:
MMEM:STOR:LIST 'test'
Stores the current list evaluation results in the test.dat file.
Manual operation: See "ASCII File Export" on page 294
See "Save Evaluation List" on page 352
See "Save Evaluation List" on page 382
See "ASCII File Export" on page 382
MMEMory:STORe<n>:PEAK
<FileName
This command exports the marker peak list to a *.dat file.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<FileName> String containing the path and name of the target file.
Example:
MMEM:STOR:PEAK 'test'
Saves the current marker peak list in the file test.dat.
Usage: Event
MMEMory:STORe:SGRam <File>
This command exports the spectrogram data to a ASCII file.
The file contains the data for every frame in the history buffer. The data corresponding
to a particular frame begins with information about the frame number and the time that
frame was recorded.
Note that, depending on the size of the history buffer, the process of exporting the data
can take a while.
Parameters:
<File> string containing the file name of the destination file
Example:
MMEM:STOR:SGR
Copies the spectrogram data to a file.
Manual operation: See "ASCII Trace Export" on page 200