[SENSe:]ESPectrum:RANGe<range>:SWEep:TIME:AUTO................................................803
[SENSe:]ESPectrum:RANGe<range>:TRANsducer...........................................................804
[SENSe:]ESPectrum:RRANge.........................................................................................804
[SENSe:]ESPectrum:RTYPe........................................................................................... 804
[SENSe:]ESPectrum:BWID <Bandwidth>
This command defines the bandwidth used for measuring the channel power (refer-
ence range). This setting takes only effect if channel power is selected as power refer-
ence type (see [SENSe:]ESPectrum:RTYPe on page 804).
Parameters:
<Bandwidth> minimum span ≤ value ≤ span of reference range
*RST: 3.84 MHz
Example:
ESP:RTYP CPOW
Sets the power reference type to channel power.
ESP:BWID 1 MHZ
Sets the Tx bandwidth to 1 MHz.
Manual operation: See "Edit Reference Range" on page 351
[SENSe:]ESPectrum:FILTer[:RRC][:STATe] <State>
This command activates or deactivates the use of an RRC filter. This setting only takes
effect if channel power is selected as power reference type (see [SENSe:
]ESPectrum:RTYPe on page 804).
Parameters:
<State> ON | OFF
*RST: ON
Example:
ESP:RTYP CPOW
Sets the power reference type to channel power.
ESP:FILT OFF
Deactivates the use of an RRC filter.
Manual operation: See "Edit Reference Range" on page 351
[SENSe:]ESPectrum:FILTer[:RRC]:ALPHa <Value>
This command sets the alpha value of the RRC filter. This setting takes only effect if
channel power is selected as power reference type ( [SENSe:]ESPectrum:RTYPe
command) and if the RRC filter is activated ( [SENSe:]ESPectrum:FILTer[:
RRC][:STATe]command).
Parameters:
<Value> 0 to 1
*RST: 0.22