EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 818

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
818Operating Manual 1176.7510.02 ─ 09
Parameters:
<Frequency> see rules in chapter 4, "Ranges and settings"
*RST: -2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Example:
LIST:RANG3:STOP 10000000
Sets the stop frequency for range 2 to 10 MHz.
Manual operation: See "Range Start / Range Stop" on page 378
[SENSe:]LIST:RANGe<range>:FILTer:TYPE
<Type>
This command sets the filter type for the specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<Type> NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
PULSe
EMI (6dB) filters
P5
5 Pole filters
*RST: NORM
The available bandwidths of the filters are specified in the data
sheet.
Example:
LIST:RANG1:FILT:TYPE RRC
Sets the RRC filter type for range 1.
Manual operation: See "Filter Type" on page 378
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation <Value>
This command sets the attenuation for the specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<Value> Refer to the data sheet.
*RST: 0 dB
Example:
LIST:RANG3:INP:ATT 10
Sets the attenuation of range 3 to 10 dB.
Remote Control – Commands