Remote Control
R&S
®
FSVA/FSV
817Operating Manual 1176.7510.02 ─ 09
Parameters:
<Detector> NEGative | POSitive | SAMPle | RMS | AVERage
NEGative
minimum peak detector
POSitive
peak detector
SAMPle
sample detector
RMS
RMS detector
AVERage
average detector
*RST:
RMS
Example:
LIST:RANGe3:DET SAMP
Sets the sample detector for range 3.
Manual operation: See "Detector" on page 379
[SENSe:]LIST:RANGe<range>[:FREQuency]:STARt <Frequency>
This command sets the start frequency for the specified range.
In order to change the start/stop frequency of the first/last range, select the appropriate
span. If you set a span that is smaller than the overall span of the ranges, the mea-
surement includes only the ranges that lie within the defined span and have a minimum
span of 20 Hz. The first and last range are adapted to the given span as long as the
minimum span of 20 Hz is not violated.
Suffix:
<range>
.
1...20
range
Parameters:
<Frequency> see rules in chapter 4, "Ranges and settings"
*RST: -250.0 MHz (range 1), -2.52 MHz (range 2), 2.52
MHz (range 3)
Example:
LIST:RANG1:STAR 100000000
Sets the start frequency for range 1 to 100 MHz.
Manual operation: See "Range Start / Range Stop" on page 378
[SENSe:]LIST:RANGe<range>[:FREQuency]:STOP <Frequency>
This command sets the stop frequency for the specified range. See also [SENSe:
]LIST:RANGe<range>[:FREQuency]:STARt on page 817.
Suffix:
<range>
.
1...20
range
Remote Control – Commands