List and sweep mode
R&S
®
SMB100B
117User Manual 1178.3711.02 ─ 08
Table 6-2: Cross-reference between manual and remote control in Single / Extern Single modes
(Sweep/List)
Manual control mode:
"Single / Extern Single"
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep :TRIGger<hw>:FSWeep:SOURce SINGle [BUS] "Single" or
:TRIGger<hw>:FSWeep:SOURce EXTernal [EXTernal] "Extern Single"
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep :TRIGger<hw>:PSWeep:SOURce SINGle [BUS] for "Single" or
:TRIGger<hw>:PSWeep:SOURce EXTernal [EXTernal] for "Extern Sin-
gle"
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep :TRIGger<hw>:LFFSweep:SOURce SINGle [BUS] "Single" or
:TRIGger<hw>:LFFSweep:SOURce EXTernal [EXTernal] "Extern Sin-
gle"
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO
List :SOURce<hw>:LIST:TRIGger:SOURce SINGle "Single" or
:SOURce<hw>:LIST:TRIGger:SOURce EXTernal "Extern Single"
:SOURce<hw>:LIST:MODE AUTO
Step / Extern Step mode (Sweep/List)
Figure 6-5: Step / Extern Step mode (sweep / list )
●
The instrument generates the signal step-by-step.
●
Starts signal generation with "State = On".
●
Switches to the next step with a manual action.
If the end value is reached, signal generation starts again from the beginning.
●
Stops signal generation with "State = Off".
Signal generation and triggering in the sweep and list modes