EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Sense, READ, Initiate and Slist Subsystems

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
396User Manual 1178.3711.02 ─ 08
Example:
:OUTPut1:AMODe FIXed
:OUTPut1:AFIXed:RANGe:UPPer?
// -27
:OUTPut1:AFIXed:RANGe:LOW?
// -50
Usage: Query only
Manual operation: See "Level Range" on page 153
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped.
To define the output state, use the command :OUTPut<hw>[:STATe].
Example:
:OUTPut1:PROT:CLE
Resets the protective circuit of the RF output.
Usage: Event
Manual operation: See "Overload" on page 153
:OUTPut<hw>:PROTection:TRIPped?
Queries the state of the protective circuit.
Return values:
<Tripped> 1 | ON | 0 | OFF
*RST: 0
Example:
:OUTPut1:PROT:TRIP
queries the state of the protective circuit of the RF output.
Response: 0
the protective circuit has not tripped.
Usage: Query only
Manual operation: See "Overload" on page 153
12.14 SENSe, READ, INITiate and SLISt subsystems
These subsystems contain the commands for configuring the power measurements
with R&S NRP power sensor connected to the R&S SMB100B.
SENSe, READ, INITiate and SLISt subsystems

Table of Contents

Other manuals for Rohde & Schwarz R&S SMB100B

Related product manuals