Remote control commands
R&S
®
SMB100B
394User Manual 1178.3711.02 ─ 08
Suffix Value range Description
OUTPut<hw>
[1] Optional suffix
:OUTPut:ALL[:STATe].....................................................................................................394
:OUTPut<hw>[:STATe]....................................................................................................394
:OUTPut<hw>[:STATe]:PON............................................................................................394
:OUTPut<hw>:AMODe................................................................................................... 395
:OUTPut<hw>:IMPedance?.............................................................................................395
:OUTPut<hw>:AFIXed:RANGe:LOWer?...........................................................................395
:OUTPut<hw>:AFIXed:RANGe:UPPer?............................................................................395
:OUTPut<hw>:PROTection:CLEar................................................................................... 396
:OUTPut<hw>:PROTection:TRIPped?..............................................................................396
:OUTPut:ALL[:STATe] <State>
Activates the RF output signal of the instrument.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: n.a. (factory preset: 0)
Example:
:OUTPut:ALL:STATe 0
:OUTPut<hw>[:STATe] <State>
Activates the RF output signal.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example:
:OUTPut1 ON
Activates the RF output.
Manual operation: See "RF State/RF ON" on page 58
:OUTPut<hw>[:STATe]:PON <Pon>
Defines the state of the RF output signal when the instrument is switched on.
Parameters:
<Pon> OFF | UNCHanged
*RST: n.a. (factory preset: UNCHanged)
Example:
:OUTPut1:PON OFF
The RF output is deactivated when the instrument is switched
on.
Manual operation: See "Power-On State" on page 230
OUTPut subsystem