Improving level performance
R&S
®
SMB100B
160User Manual 1178.3711.02 ─ 08
State
Activates user correction.
The R&S SMB100B displays the status icon "Lev Ucor" in the "Level" panel.
Remote command:
[:SOURce<hw>]:CORRection[:STATe] on page 451
User Correction
Indicates the corrected level value for a specific frequency point.
Remote command:
[:SOURce<hw>]:CORRection:VALue? on page 450
UCOR Data
Accesses the standard "File Select" function of the instrument. The provided navigation
possibilities in the dialog are self-explanatory.
Files with user correction values are files with predefined file extension *.uco. When a
file is selected, the dialog indicates the filename.
You can create the file internally in the table editor or externally.
●
To select an existing file, select "Select List > navigate to the file *.uco > Select"
●
Use the general editor function to create internally new file or to edit an existing
one.
●
Use the standard file manager function to load externally created files to the instru-
ment.
Remote command:
[:SOURce]:CORRection:CSET:CATalog? on page 451
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 450
[:SOURce]:CORRection:CSET:DELete on page 451
Edit UCOR Data
Opens the build-in table editor to define a new correction table or edit an existing one.
See also:
●
Chapter 6.7, "List editor", on page 145
●
"Fill..." on page 93
●
Chapter 7.3.3, "Fill with sensor", on page 164
UCOR Info ...
Opens a window information on the user correction file.
The window shows the metadata of the file and the output level parameters that are
relevant for user correction.
Remote command:
n.a.
7.3.2 List editor
The "User Correction" and "List Mode" dialogs provide a build-in list editor for defining
the corresponding value pairs.
User correction