To initiate a ping at the instrument:
1. On the "Ping Client" page, enter the IP address of the host in the "Destination
Address" field (for example 10.113.1.151).
2. Select "Submit".
11.6.2.2 SCPI remote trace
The remote trace functionality allows you to trace input and output strings at the
remote control interface of the R&S SMB100B, see Chapter 11.11, "To trace messages
with the LXI web browser interface", on page 329.
A recorded trace (message log) can be evaluated directly in the dialog. Use the high-
lighting and navigation functions provided by the lower toolbar to locate error mes-
sages and messages containing arbitrary search strings. You can also export the mes-
sage log to a *.csv file and evaluate the file using a suitable program.
To trace and display messages, switch on "logging" and "live mode" in the toolbar.