Remote control commands
R&S
®
SMB100B
449User Manual 1178.3711.02 ─ 08
12.15.2.1 Correction settings
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency <Frequency>
Enters the frequency value in the table selected with [:SOURce<hw>]:CORRection:
CSET[:SELect].
Parameters:
<Frequency> Frequency#1[, Frequency#2, ...]
String of values with default unit Hz.
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Manual operation: See "Edit List Mode Data" on page 146
[:SOURce<hw>]:CORRection:CSET:DATA:POWer <Power>
Enters the level values to the table selected with [:SOURce<hw>]:CORRection:
CSET[:SELect].
Parameters:
<Power> Power#1[, Power#2, ...]
String of values with default unit dB.
*RST: 0
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Manual operation: See "Edit List Mode Data" on page 146
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of frequency/level values in the selected table.
Return values:
<Points> integer
Range: 0 to 10000
*RST: 0
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Usage: Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Fills the selected user correction table with the level values measured by the power
sensor for the given frequencies.
To select the used power sensor set the suffix in key word SENSe.