EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Source:frequency Subsystem

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
454User Manual 1178.3711.02 ─ 08
Parameters:
<Filename> string
Filename or complete file path; file extension can be omitted.
Example: See Example"User correction data exchange" on page 448 .
Manual operation: See "Select Source/Select ASCII Destination" on page 96
12.15.3 SOURce:FREQuency subsystem
The SOURce:FREQuency subsystem contains the commands used to define the fre-
quency settings for the RF sources and sweeps.
Example: Frequency configuration
SOURce1:FREQuency:MODE CW
SOURce1:FREQuency:CW 6000000000
SOURce1:FREQuency:OFFSet 2000000000
SOURce1:FREQuency:MULTiplier 1.5
SOURce1:FREQuency:CW?
// 11000000000
// SOURce1:FREQuency:STEP:MODE USER
// SOURce1:FREQuency:STEP:INCRement 1000000
// SOURce1:FREQuency:CW UP
SOURce1:PHASe 2
SOURce1:PHASe:REFerence
[:SOURce<hw>]:FREQuency:MODE................................................................................454
[:SOURce<hw>]:FREQuency[:CW|FIXed].........................................................................455
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL................................................................. 456
[:SOURce<hw>]:FREQuency:MANual..............................................................................457
[:SOURce<hw>]:FREQuency:MULTiplier.......................................................................... 457
[:SOURce<hw>]:FREQuency:OFFSet.............................................................................. 457
[:SOURce<hw>]:FREQuency:CENTer..............................................................................458
[:SOURce<hw>]:FREQuency:SPAN.................................................................................458
[:SOURce<hw>]:FREQuency:STARt................................................................................ 458
[:SOURce<hw>]:FREQuency:STOP.................................................................................459
[:SOURce<hw>]:FREQuency:STEP:MODE...................................................................... 459
[:SOURce<hw>]:FREQuency:STEP[:INCRement]............................................................. 460
[:SOURce<hw>]:FREQuency:MODE <Mode>
Sets the frequency mode for generating the RF output signal. The selected mode
determines the parameters to be used for further frequency settings.
Parameters:
<Mode> CW | FIXed | SWEep | LIST | COMBined
SOURce subsystem

Table of Contents

Other manuals for Rohde & Schwarz R&S SMB100B

Related product manuals