Improving level performance
R&S
®
SMB100B
164User Manual 1178.3711.02 ─ 08
"From / Range"
Defines the start line and number of lines to be filled.
"Column To Fill"
Selects, if the column is filled up with frequencies in Hz, levels in dBm
or dwell times in s.
"Start Value / End Value"
Sets the start value for frequency, level or dwell time. The end value
is read only and depends on the increment value and the range.
"Increment Value"
Determines the step size.
"Fill"
Fills the column specified in "Column To fill".
7.3.3 Fill with sensor
Access:
1. Select "Level" > "User Correction".
2. Select "UCOR Data > navigate to the file *.ucor > Select".
3. Select "Edit UCOR Data > Fill With Sensor...".
This dialog contains parameters for filling a table automatically with sensor read-
ings.
User correction