EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - List Mode File Operation

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
482User Manual 1178.3711.02 ─ 08
Rohde & Schwarz parameter SCPI parameter Applies to the list mode parameters:
AUTO IMMediate [:SOURce<hw>]:LIST:MODE AUTO
SINGle BUS
[:SOURce<hw>]:LIST:MODE AUTO or
[:SOURce<hw>]:LIST:MODE STEP
EXTernal EXTernal
[:SOURce<hw>]:LIST:MODE AUTO or
[:SOURce<hw>]:LIST:MODE STEP
Parameters:
<Source> AUTO | IMMediate | SINGle | BUS | EXTernal
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled
continuously. The selected list is restarted as soon as it is fin-
ished.
SINGle|BUS
The list is triggered by the command [:SOURce<hw>]:LIST:
TRIGger:EXECute. The list is executed once.
EXTernal
The list is triggered externally and executed once.
*RST: AUTO
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Manual operation: See "Mode" on page 140
[:SOURce<hw>]:LIST:RUNNing?
Queries the current state of the list mode.
Return values:
<State> 1 | ON | 0 | OFF
1
Signal generation based on the list mode is active.
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Usage: Query only
12.15.6.2 List mode file operation
The following section covers basic commands to file handling in list mode.
[:SOURce<hw>]:LIST:CATalog?.......................................................................................483
[:SOURce<hw>]:LIST:DELete..........................................................................................483
[:SOURce<hw>]:LIST:DELete:ALL................................................................................... 483
SOURce subsystem