Remote control commands
R&S
®
SMB100B
481User Manual 1178.3711.02 ─ 08
You can either enter the data as a list of numbers, or as binary
block data.
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also :FORMat[:DATA].
Range: depends on the installed options
Default unit: dBm
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Manual operation: See "Edit List Mode Data" on page 146
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the number (points) of level entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Usage: Query only
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode.
Example:
See Example"Create an RF list and activate the list mode"
on page 474.
Usage: Event
Manual operation: See "Execute Single" on page 141
[:SOURce<hw>]:LIST:TRIGger:SOURce <Source>
Selects the trigger source for processing lists.
The designation of the parameters correspond to those in sweep mode. SCPI standard
uses other designations for the parameters, which are also accepted by the instrument.
The SCPI designation should be used if compatibility is an important consideration. For
an overview, see the following table:
SOURce subsystem