List and sweep mode
R&S
®
SMB100B
136User Manual 1178.3711.02 ─ 08
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing on page 511
RF combined sweep: n.a.
LF frequency sweep:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing on page 473
Step Linear/Step Logarithmic
In "RF/LF Frequency Sweep" mode, sets the step width for the individual frequency
sweep steps. The value is added at each sweep step to the current frequency.
Depending on the current Spacing, you can enter either an absolute or logarithmic step
width.
See Chapter 6.2.1, "Correlating parameters in sweep mode", on page 122.
"Step Linear"
The step width is a constant value in Hz.
"Step Logarithmic"
The step width is determined logarithmically in %, i.e. as a constant
fraction of the current frequency.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
on page 513
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
on page 513
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:
LINear] on page 474
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:
LOGarithmic on page 473
6.5.3 Level range settings
Access:
1. Select "Sweep" > "RF Level Sweep"
2. Select for example "Level Range".
Table 6-8: Range settings dialog: Examples of RF Level and RF Combined Sweep dialogs
> "Level Range" > "Frequency / Level Range"
Sweep mode settings

Table of Contents