Remote control commands
R&S
®
SMB100B
393User Manual 1178.3711.02 ─ 08
:HCOPy:FILE[:NAME]:AUTO[:FILE]:PREFix <Prefix>
:HCOPy:FILE[:NAME]:AUTO[:FILE]:PREFix:STATe <State>
Uses the prefix for the automatic generation of the file name, provided PREF:STAT is
activated.
Parameters:
<State> 1 | ON | 0 | OFF
*RST:  1
Example:  See Example"Store a hard copy of the display" on page 388
Manual operation:  See "Prefix, Year, Month, Day" on page 224
12.12
 KBOard subsystem
The KBOard subsystem contains the commands to set a connected keyboard.
:KBOard:LAYout.............................................................................................................393
:KBOard:LAYout <Layout>
Selects the language for an external keyboard and assigns the keys acccordingly.
Parameters:
<Layout> CHINese | DANish | DUTCh | DUTBe | ENGLish | ENGUK |
FINNish | FRENch | FREBe | FRECa | GERMan | ITALian |
JAPanese | KORean | NORWegian | PORTuguese | RUSSian |
SPANish | SWEDish | ENGUS
*RST:  n.a. (factory preset: ENGLish)
Example: 
:KBOard:LAYout US
// activates American keyboard
Manual operation:  See "USB Keyboard > Layout" on page 229
12.13  OUTPut subsystem
In the OUTPut subsystem, you can configure the output signals.
The LF output signal is defined with the commands of the Chapter 12.15.5,
"SOURce:LFOutput subsystem", on page 461 system.
Common suffixes
The following common suffixes are used in remote commands:
OUTPut subsystem