General instrument functions
R&S
®
SMB100B
230User Manual 1178.3711.02 ─ 08
This dialog enables you, to deactivate updating the display.
The remote command to switch off the display update is described in Chapter 12.9,
"DISPlay subsystem", on page 382.
Display Update is
Disables the automatic refreshing of the displayed values.
Remote command:
:DISPlay:UPDate[:STATe] on page 384
10.1.3 Defining the RF signal state on power on
As additional functions for the level settings, you can determine the state of the RF sig-
nal, and the level display in the status bar when you turn on the R&S SMB100B.
Power-On/EMF settings
Access:
Select "Level" > "Power-On / EMF".
The "Power-On/EMF" dialog contains all settings for configuring the power-on
behavior and the level display.
Settings:
Power-On State...........................................................................................................230
Display Level as Voltage of EMF................................................................................ 230
Power-On State
Determines the RF signal output state when the instrument is switched on.
You can disable the RF output signal in general, or start it in the same state as it had
been when it was switched off.
Remote command:
:OUTPut<hw>[:STATe]:PON on page 394
Display Level as Voltage of EMF
Activates display of the signal level as voltage of the EMF (no-load voltage).
If disabled, the level is displayed as a voltage over a 50 Ohm load.
Note: