Improving level performance
R&S
®
SMB100B
180User Manual 1178.3711.02 ─ 08
Each log file contains the measured value (2 readings when you work with peak sen-
sors), the sensor type, and the measurement time (timestamp). Logged data is not
overwritten. When a new measurement is started, the collected logging data is appen-
ded in the log file.
Check the used disc space regularly and remove log files to maintain storage capacity.
Note: The logging function is intended for measurements with long time intervals. It is
suitable source for data reconstructions if the connection to the sensor was interrupted.
Remote command:
:SENSe<ch>[:POWer]:LOGGing:STATe on page 409
7.5 How to calibrate the power level with an R&S NRP
power sensor
Using a R&S NRP power sensor and the user correction function of the instrument,
you can compensate external frequency responses. The R&S SMB100B utilizes the
readings of the power sensor and creates a correction value table for controlling the
output level during operation.
To connect the R&S NRP to the R&S SMB100B
1. Connect the power sensor.
See Chapter 7.4.1, "Connecting R&S NRP power sensors to the R&S SMB100B",
on page 168.
2. Select "Power Sens" > "NRP Sensor Mapping".
3. If the sensor is not detected automatically, select "Scan > Start".
The instrument scans the network and the USB connections for connected sensors
and lists all detected R&S NRP sensors in the mapping table.
4. In the "Mapping" column, assign the sensor to a sensor channel, see Chap-
ter 7.4.2, "NRP sensor mapping", on page 169.
5. Close the dialog.
To configure and calibrate the R&S NRP in the R&S SMB100B
Provided the power sensor is connected to the R&S SMB100B and is assigned to a
sensor channel, we recommend that you calibrate and configure the power sensor in
the "NRP Power Viewer" dialog.
1. Select "Level" > "RF ON > Off"
2. Select "Power Sens" > "NRP Power Viewer".
How to calibrate the power level with an R&S NRP power sensor