Remote control commands
R&S
®
SMB100B
409User Manual 1178.3711.02 ─ 08
:SENSe<ch>[:POWer]:LOGGing:STATe <State>
Activates the recording of the power values, measured by a connected R&S NRP
power sensor.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example:
:SENS:LOGG:STAT ON
Activates recording of the power measurement of the first sen-
sor.
Manual operation: See "Enable Logging" on page 179
:SENSe<ch>[:POWer]:OFFSet <Offset>
Sets a level offset which is added to the measured level value after activation with
command :SENSe<ch>[:POWer]:OFFSet:STATe. The level offset allows, e.g. to
consider an attenuator in the signal path.
Parameters:
<Offset> float
Range: -100.0 to 100.0
*RST: 0
Default unit: dB
Example:
:SENS1:POW:OFFS 10.0
Sets a level offset of 10 dB
Manual operation: See "Level Offset State,Level Offset" on page 177
:SENSe<ch>[:POWer]:OFFSet:STATe <State>
Activates the addition of the level offset to the measured value. The level offset value is
set with command :SENSe<ch>[:POWer]:OFFSet.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example:
:SENS1:POW:OFFS 0.4dB
Sets a level offset of 0.4 dB
:SENS1:POW:OFFS:STAT ON
A level offset of 0.4 dB is added to the measured value.
Manual operation: See "Level Offset State,Level Offset" on page 177
:SENSe<ch>[:POWer]:SNUMber?
Queries the serial number of the sensor.