EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Page 408

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
408User Manual 1178.3711.02 ─ 08
Manual operation: See "Auto Once" on page 178
:SENSe<ch>[:POWer]:FILTer:TYPE <Type>
Selects the filter mode. The filter length is the multiplier for the time window and thus
directly affects the measurement time.
Parameters:
<Type> AUTO | USER | NSRatio
AUTO
Automatically selects the filter length, depending on the mea-
sured value. The higher the power, the shorter the filter length,
and vice versa.
USER
Allows you to set the filter length manually. As the filter-length
takes effect as a multiplier of the measurement time, you can
achieve constant measurement times.
NSRatio
Selects the filter length (averaging factor) according to the crite-
rion that the intrinsic noise of the sensor (2 standard deviations)
does not exceed the specified noise content. You can define the
noise content with command :SENSe<ch>[:POWer]:FILTer:
NSRatio.
Note: To avoid long settling times when the power is low, you
can limit the averaging factor limited with the "timeout" parame-
ter (:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe).
*RST:
AUTO
Example: See Example"Performing a power measurement with a fixed fil-
ter" on page 399.
Manual operation: See "Filter" on page 178
:SENSe<ch>[:POWer]:FREQuency <Frequency>
Sets the RF frequency of the signal, if signal source :SENSe<ch>[:POWer]:SOURce
USER is selected.
Parameters:
<Frequency> float
*RST: 1 GHz
Example:
:SENS1:SOUR USER
Selects user-defined source.
:SENS1:FREQ 2.44GHz
Sets the RF frequency of the source which is 2.44 GHz.
Manual operation: See "Frequency" on page 177
SENSe, READ, INITiate and SLISt subsystems

Table of Contents

Other manuals for Rohde & Schwarz R&S SMB100B

Related product manuals