Remote control commands
R&S
®
SMB100B
371User Manual 1178.3711.02 ─ 08
If you have enabled automatic shutdown, :CALibration:DELay:SHUTdown[:
STATe] ON, the instrument shuts down when the adjustments are completed.
Return values: 
<Error> 1 | ON | 0 | OFF
*RST:  n.a. (no preset. default: 0)
Example: 
:CALibration:DELay[:MEASure] ON|1
// enables the adjustment processs to start after the warm-up time automatically
Usage:  Query only
Manual operation:  See "Adjust All Delayed" on page 590
12.7  DIAGnostic subsystem
The DIAGnostic subsystem contains the commands used for instrument diagnosis
and servicing. SCPI does not define any DIAGnostic commands; the commands lis-
ted here are all device-specific. All DIAGnostic commands are query commands
which are not affected by *RST.
The test functions are intended for services purposes.
They are thus password-protected functions. Unlock the corresponding protection level
to access them, see :SYSTem:PROTect<ch>[:STATe].
For more information, see R&S SMB100B Service Manual.
Common suffixes
The following common suffixes are used in remote commands:
Suffix Value range Description
DIAGnostic<hw>
[1] Optional suffix
DIAGnostic subsystem