Maintenance, storage and disposal
R&S
®
SMB100B
590User Manual 1178.3711.02 ─ 08
Settings
Adjust All Delayed.......................................................................................................590
Warm Up Time............................................................................................................ 590
Shutdown After Adjustment.........................................................................................590
Adjust All Delayed
Performs all available internal calibration routines of the instrument.
How to: See Starting the internal adjustments automatically.
Remote command:
:CALibration:DELay[:MEASure]? on page 370
Warm Up Time
Sets the time for warming up the instrument before the calibration starts automatically.
How to: See Starting the internal adjustments automatically.
Remote command:
:CALibration:DELay:MINutes on page 370
Shutdown After Adjustment
Enables the automatic shutdown after internal adjustments.
How to: See Starting the internal adjustments automatically.
Remote command:
:CALibration:DELay:SHUTdown[:STATe] on page 370
15.3.1.3 How to use the internal adjustments
Deciding whether to run internal adjustments
1. Select "System Config > Setup > Internal Adjustment".
2. Observe the status and color indication in the section "Since Last Full Adjustment".
Green: internal adjustments are not required.
Red: internal adjustments are required. Observe also the indication in the "Informa-
tion" field.
3. We recommend that you run internal adjustments in the following cases:
● Before starting any application that requires a maximum of level accuracy.
● When a long period of time has passed since the last adjustments.
● If the ambient temperature of the instrument significantly differs from the one of
the last adjustments.
4. Proceed as described in "Running internal adjustments" on page 590.
Running internal adjustments
1.
NOTICE! Adjustments can be invalid if performed when the instrument is not
warmed-up.
Performing maintenance tasks