Maintenance, storage and disposal
R&S
®
SMB100B
589User Manual 1178.3711.02 ─ 08
Remote command:
:CALibration<hw>:ALL:TIME? on page 369
Temperature Offset
Displays the temperature difference, comparing the temperature of the last adjustment
to the current instrument temperature.
A green checkmark indícates that the offset is within the permitted range. If the temper-
ature deviates more than ±5 K, the instrument indicates a warning icon.
Remote command:
:CALibration<hw>:ALL:TEMP? on page 369
Information
Displays information to the current adjustment state.
Remote command:
:CALibration<hw>:ALL:INFormation? on page 368
Continue Adjustment on Error
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
This function is password-protected. Unlock the protection level 1 to access it.
Remote command:
:CALibration<hw>:CONTinueonerror on page 370
15.3.1.2 Delayed adjustment settings
Access:
► Select "System Config > Setup > General > Delayed Adjustment".
In this dialog, you can set the internal adjustments to start automatically after the
selected warm up time and to switch off the instrument after calibration.
The remote commands required to define these settings are described in Chapter 12.6,
"CALibration subsystem", on page 367.
Performing maintenance tasks