Remote control commands
R&S
®
SMB100B
369User Manual 1178.3711.02 ─ 08
Example:
CAL:ALL:INF?
"Instrument is calibrated, no adjustment required."
"UNCAL, instrument is warming up."
"UNCAL, Please perform full adjustment after warming up."
"UNCAL, Please perform full adjustment."
Usage: Query only
Manual operation: See "Information" on page 589
:CALibration<hw>:ALL:TEMP?
Queries the temperature deviation compared to the calibration temperature.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Temperature> string
Example:
CALibration:ALL:TEMP?
// "+12.00 K"
Usage: Query only
Manual operation: See "Temperature Offset" on page 589
:CALibration<hw>:ALL:TIME?
Queries the time elapsed since the last full adjustment.
Return values:
<Time> string
Example:
CAL:ALL:TIME?
// "22 days"
Usage: Query only
Manual operation: See "Time" on page 588
:CALibration:DATA:FACTory:DATE?
Queries the date of the last factory calibration.
Return values:
<Date> string
Example:
CAL:DATA:FACT:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Factory Calibration" on page 580
CALibration subsystem