Remote control commands
R&S
®
SMB100B
370User Manual 1178.3711.02 ─ 08
:CALibration<hw>:CONTinueonerror <State>
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
Suffix:
<hw>
.
[1]
Optional suffix
Parameters:
<State> 1 | ON | 0 | OFF
*RST: n.a. (factory preset: 0)
Example:
CAL:CONT ON
// Continues calibration after an error
Manual operation: See "Continue Adjustment on Error" on page 589
:CALibration:DELay:MINutes <Minutes>
Sets the warm-up time to wait before internal adjustment starts automatically.
Automatic execution starts only, if you have enabled the calibration with command :
CALibration:DELay[:MEASure]? ON.
Parameters:
<Minutes> integer
Range: 30 to 120
*RST: n.a. (no preset. default: 60)
Example:
:CALibration:DELay:MINutes 30
// sets the time delay for warm-up of the instrument.
Manual operation: See "Warm Up Time" on page 590
:CALibration:DELay:SHUTdown[:STATe] <Shutdown>
Enables the instrument to shut down automatically after calibration.
Parameters:
<Shutdown> 1 | ON | 0 | OFF
*RST: 0
Example:
:CALibration:DELay:SHUTdown[:STATe] ON|1
// inititates that the instrument shuts down when the adjustments are completed.
Manual operation: See "Shutdown After Adjustment" on page 590
:CALibration:DELay[:MEASure]?
Starts the delayed adjustment process. When the warm-up time has elapsed (see :
CALibration:DELay:MINutes, it executes the internal adjustments.
CALibration subsystem