List and sweep mode
R&S
®
SMB100B
130User Manual 1178.3711.02 ─ 08
State (RF frequency / level sweep)
Activates the RF frequency / level sweep signal generation.
Note: Active RF frequency, RF level or RF combined sweep modes deactivate the LF
sweep or List mode and vice versa.
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 454
[:SOURce<hw>]:POWer:MODE on page 495
State (LF frequency sweep)
Status of the LF frequency sweep signal.
You find this parameter in two dialogs:
●
"LF Frequency Sweep" turns on/off the signal generation.
See Sweep mode settings for access.
●
"Analog Modulation Sources" displays the current state (read-only). If it is turned
on, this dialog provides the varying sweep frequency, see Current Frequency.
Note: Active RF frequency, RF level or RF combined sweep modes deactivate the LF
sweep or List mode and vice versa.
Remote command:
[:SOURce<hw>]:LFOutput:FREQuency:MODE on page 465
Current Frequency
Displays the current frequency sweep value for all sweep Modes, except of Mode >
Step. In this mode, you can enter the next frequency setting manually.
Remote command:
[:SOURce<hw>]:FREQuency:MANual on page 457
[:SOURce<hw>]:LFOutput:FREQuency:MANual on page 465
Current Level
Displays the current level sweep value for all sweep Modes, except of Mode > Step. In
this mode, you can enter the next level setting manually.
Remote command:
[:SOURce<hw>]:POWer:MANual on page 495
Mode
Selects the sweep mode.
See Chapter 6.1, "Signal generation and triggering in the sweep and list modes",
on page 114.
"Auto"
Generates a continuously repeating sweep signal directly after acti-
vating the sweep mode.
The sweep steps are performed automatically, controlled by the dwell
time.
"Single / Extern Single"
Generates a single sweep cycle after a trigger event.
The sweep steps within the cycle are performed automatically, con-
trolled by the dwell time. If one cycle is completed, the instrument
waits for the next trigger event.
Sweep mode settings