Remote control commands
R&S
®
SMB100B
495User Manual 1178.3711.02 ─ 08
Parameters:
<Amplitude> float
Range: depends on the installed options
Increment: 0.01
*RST: n.a. (factory preset: 30)
Example:
SOURce1:POWer:LIMit:AMPLitude 10
Manual operation: See "Limit" on page 64
[:SOURce<hw>]:POWer:MANual <Manual>
Sets the level for the subsequent sweep step if SWE:POW:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any level within the setting range, where:
STARt is set with [:SOURce<hw>]:POWer:STARt
STOP is set with [:SOURce<hw>]:POWer:STOP
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate]:OFFSet
Range: (STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01
Default unit: dBm
Example: See Example"Setup an RF frequency or power sweep"
on page 506
Manual operation: See "Current Level" on page 130
[:SOURce<hw>]:POWer:MODE <Mode>
Selects the operating mode of the instrument to set the output level.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW and FIXed are synonyms.
To set the output level value, use the command [:
SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude].
SWEep
Sets sweep mode.
Set the range and current level with the commands:
[:SOURce<hw>]:POWer:STARt and [:SOURce<hw>]:
POWer:STOP,
[:SOURce<hw>]:POWer:MANual.
SOURce subsystem