RF signal configuration
R&S
®
SMB100B
64User Manual 1178.3711.02 ─ 08
Settings
RF State/RF ON............................................................................................................64
Amplitude...................................................................................................................... 64
Limit...............................................................................................................................64
Offset.............................................................................................................................65
Setting Characteristics.................................................................................................. 65
Level Range..................................................................................................................66
Readjust........................................................................................................................66
User Variation................................................................................................................66
└ Variation Active............................................................................................... 66
└ Variation Step..................................................................................................66
RF State/RF ON
Activates or deactivates the RF output.
Acts as the [RF on/off] key.
Remote command:
:OUTPut<hw>[:STATe] on page 394
Amplitude
Sets the level of the RF signal.
The value is offset-free and corresponds to the level at the RF 50 Ω connector.
See "RF frequency and level display with a downstream instrument" on page 59.
Note: Suppressed values in the status bar
For security concerns or certain operating modes, you can hide the frequency and
level display in the status bar.
●
The display has been disabled for security reasons.
See:
– Annotation Frequency
– Annotation Amplitude
●
The display is disabled when list mode is running, see Chapter 6, "List and sweep
mode", on page 112.
Remote command:
[:SOURce<hw>]:POWer:POWer on page 496
Note: The SCPI command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude] sets the level of the "Level" display.
This means, the level containing offset.
Limit
Sets an upper limit for the RF output power.
You can use this value to protect your DUT from damage due to high input power. If
you enter an RF level above this value, the instrument limits the output power to this
specified value, and generates the warning message:
RF level settings