List and sweep mode
R&S
®
SMB100B
128User Manual 1178.3711.02 ─ 08
Live list processing mode
The R&S SMB100B generates the signal directly from the value pairs in the database,
and adjusts the hardware settings accordingly. The current instrument state and thus
any change during the signal generation directly affects the RF signal. The temporary
memory is not used.
You can conveniently modify parameters like modulation settings during run-time.
Learning list mode data is not required. Impacts like temperature drift are also consid-
ered immediately.
This mode is optimized for maximum signal quality, and is useful if dwell times higher
than 2 ms are sufficient.
6.5 Sweep mode settings
This section lists the settings of all available sweep modes.
Access:
1. Select "Sweep" > "RF Frequency Sweep"
2. Select "Sweep" > "RF Level Sweep"
3. Select "Sweep" > "RF Combined Sweep"
4. Select "Sweep" > "LF Frequency Sweep"
The remote commands required to define these settings are described in:
●
Chapter 12.15.12, "SOURce:SWEep subsystem", on page 505
●
Chapter 12.15.3, "SOURce:FREQuency subsystem", on page 454
●
Chapter 12.15.10, "SOURce:POWer subsystem", on page 491
●
Chapter 12.15.5, "SOURce:LFOutput subsystem", on page 461
Settings
● General sweep settings.........................................................................................128
● Frequency range settings......................................................................................134
● Level range settings..............................................................................................136
6.5.1 General sweep settings
Access:
► Select for example "Sweep" > "RF Frequency Sweep".
Sweep mode settings