File and data management
R&S
®
SMB100B
205User Manual 1178.3711.02 ─ 08
Recall
Restores the selected configuration.
During recall, the instrument considers all related settings, for example sweeps in
active state or lists. An error message indicates the settings which cannot be imple-
mented.
Remote command:
:MMEMory:LOAD:STATe on page 365
Recall Immediate x
Loads the selected configuration from one of the intermediate memories. A message
appears if no instrument configuration is stored in this memory.
Remote command:
*RCL on page 354
File Manager
Accesses the "File Manager" dialog, see Chapter 9.8, "Using the file manager",
on page 209.
9.4.2 How to save and recall instrument settings
Instrument settings can be saved to a file and loaded again later, so that you can
repeat the tests with the same settings.
To access and recall instrument setups quickly
► Assign the appropriate action to the [★ (User)] key.
See Chapter 10.2.4, "How to assign actions to the [★ (User)] key", on page 238.
To save and recall instrument settings
1. Select "System Config > Save/Recall" > "Operation Mode > Save".
2. Select "Save Immediate 1".
The instrument saves its settings in the intermediate memory 1. The filename and
the storage location cannot be changed.
3. Adapt the instrument settings as required. Select "Save Immediate 2"
4. To restore the settings, select the "Operation Mode > Recall"
5. Select "Recall Immediate 1"
The instrument is restored to the previous state.
6. Select "Recall Immediate 2" to switch to the settings saved in the second file.
To save complete instrument settings
1. Select "System Config > Save/Recall" > "Operation Mode > Save".
2. In the file selection dialog, select a filename and storage location for the settings
file.
Saving and recalling instrument settings